Database last updated on Wed, 09 Sep 2026 07:11:01 +0000 by localhost
Critical warnings from b-files' comments for days between 2020006 and 2020006 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2020 005 05 43 45 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.538020 Volts. | 16 | 3 | B00620.191 | CARE | #191 | 05:37:31 | | 1 | -1 |
| rl: | 2020 006 18 21 20 PMT counter failure | 1 | 1 | B00620.191 | CARE | #191 | 05:38:08 | | 1 | -1 |
Instrument constant changes for the week before 2020006 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 23 rows)
HG burned out? (AP results when HG is ON) (0 sec, 9 rows)
Non-critical warnings from b-files' comments for days between 2020006 and 2020006 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2020 005 05 42 44 Reset Warm -> Tepid from 2020 005 05 42 38 until 2020 005 05 42 39 . Counts: 1,0 | 1 | 3 | B00620.191 | CARE | #191 | 05:37:29 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 2 | 5 | B00620.191 | CARE | #191 | 17:45:13 | | 1 | -1 |
| ds: | DS intensity too high ( 287586/cy) quitting. | 3 | 99 | B00620.191 | CARE | #191 | 18:21:33 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
| 006 | CARE | #191 | ecnouv | 20 | 22 | 1 | 1 | 36 | 6 | 11 | 11 | 1 | 7 | 68 |
(0 sec)
Relative and absolute humidity for day 2020006 (3 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 6.7 | 1.1 |