Database last updated on Wed, 09 Sep 2026 07:11:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2020006 and 2020006 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 005 05 43 45 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.538020 Volts.163B00620.191CARE#19105:37:311-1
rl:2020 006 18 21 20 PMT counter failure11B00620.191CARE#19105:38:081-1

Instrument constant changes for the week before 2020006 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 23 rows)

HG burned out? (AP results when HG is ON) (0 sec, 9 rows)

Non-critical warnings from b-files' comments for days between 2020006 and 2020006 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 005 05 42 44 Reset Warm -> Tepid from 2020 005 05 42 38 until 2020 005 05 42 39 . Counts: 1,013B00620.191CARE#19105:37:291-1
ds:Suppressed oscillation of filter wheel.25B00620.191CARE#19117:45:131-1
ds:DS intensity too high ( 287586/cy) quitting.399B00620.191CARE#19118:21:331-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprsslzs
006CARE#191ecnouv20221136611111768
(0 sec)

Relative and absolute humidity for day 2020006 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1916.71.1