Database last updated on Wed, 09 Sep 2026 11:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2019345 and 2019345 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 344 05 23 54 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.500382 Volts.163B34519.191CARE#19105:26:051-1
rl:2019 344 06 15 40 PMT counter failure11B34519.191CARE#19105:26:161-1

Instrument constant changes for the week before 2019345 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
191201934534.8

Non-critical warnings from b-files' comments for days between 2019345 and 2019345 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 344 05 22 52 Reset Warm -> Tepid from 2019 344 05 22 47 until 2019 344 05 22 47 . Counts: 1,013B34519.191CARE#19105:26:031-1
ds:DS intensity too high ( 238464/cy) quitting.499B34519.191CARE#19116:02:091-1
ds:Suppressed oscillation of filter wheel.15B34519.191CARE#19116:58:501-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprsslzs
345CARE#191ecnouv37391148814121650
(0 sec)

New operators? (0 sec, 88 rows)

OperatorSerial
Time191
(CUT)191
(HH:MM)191

Relative and absolute humidity for day 2019345 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1915.20.9