Database last updated on Sat, 12 Sep 2026 14:40:53 +0000 by localhost
Critical warnings from b-files' comments for days between 2019302 and 2019302 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 301 05 15 05 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.699324 Volts. | 36 | 3 | B30219.191 | CARE | #191 | 05:08:01 | | 1 | -1 |
Instrument constant changes for the week before 2019302 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 32 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 191 | 2019302 | 2 | 3.8 |
Non-critical warnings from b-files' comments for days between 2019302 and 2019302 (1 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 301 05 13 49 Reset Warm -> Tepid from 2019 301 05 13 43 until 2019 301 05 13 44 . Counts: 1,0 | 1 | 3 | B30219.191 | CARE | #191 | 05:07:58 | | 1 | -1 |
| ds: | DS intensity too high ( 200554/cy) quitting. | 1 | 99 | B30219.191 | CARE | #191 | 14:01:47 | | 1 | -1 |
| ab_sk: | HOME key pressed | 1 | 99 | B30219.191 | CARE | #191 | 17:59:30 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | zs |
| 302 | CARE | #191 | ecnouv | 38 | 39 | 1 | 1 | 7 | 16 | 13 | 1 | 8 | 59 |
(1 sec)
Relative and absolute humidity for day 2019302 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 16.1 | 2.8 |