Database last updated on Sat, 12 Sep 2026 20:11:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2019154 and 2019154 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 153 05 30 37 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.183151 Volts.73B15419.191CARE#19105:24:061-1
rl:2019 154 13 01 02 PMT counter failure11B15419.191CARE#19105:24:151-1
ds:Suppressed oscillation of filter wheel.55B15419.191CARE#19113:56:301-1

Instrument constant changes for the week before 2019154 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 16 rows)

Non-critical warnings from b-files' comments for days between 2019154 and 2019154 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2019 153 05 29 21 Reset Warm -> Tepid from 2019 153 05 29 15 until 2019 153 05 29 16 . Counts: 1,013B15419.191CARE#19105:24:041-1
rl:2019 153 05 30 37 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.183151 Volts.43B15419.191CARE#19105:24:061-1
rl:2019 154 16 13 18 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.059484 Volts.33B15419.191CARE#19105:24:171-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprsslzs
154CARE#191ecnouv145149111215141732
(0 sec)

Relative and absolute humidity for day 2019154 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1916.41.2