Database last updated on Sat, 12 Sep 2026 16:40:53 +0000 by localhost
Critical warnings from b-files' comments for days between 2019143 and 2019143 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 142 05 30 29 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.645555 Volts. | 29 | 3 | B14319.191 | CARE | #191 | 05:21:59 | | 1 | -1 |
| rl: | 2019 143 16 14 34 PMT counter failure | 1 | 1 | B14319.191 | CARE | #191 | 05:22:59 | | 1 | -1 |
Instrument constant changes for the week before 2019143 (4 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 30 rows)
Non-critical warnings from b-files' comments for days between 2019143 and 2019143 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 142 05 29 13 Reset Warm -> Tepid from 2019 142 05 29 07 until 2019 142 05 29 08 . Counts: 1,0 | 1 | 3 | B14319.191 | CARE | #191 | 05:21:56 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 4 | 5 | B14319.191 | CARE | #191 | 16:34:16 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
| 143 | CARE | #191 | ecnouv | 81 | 88 | 1 | 1 | 8 | 7 | 20 | 18 | 1 | 7 | 69 |
(1 sec)
Relative and absolute humidity for day 2019143 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 15.2 | 3.4 |