Database last updated on Sun, 13 Sep 2026 01:41:14 +0000 by localhost
Critical warnings from b-files' comments for days between 2019127 and 2019127 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 126 05 27 24 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.543397 Volts. | 25 | 3 | B12719.191 | CARE | #191 | 05:20:11 | | 1 | -1 |
| rl: | 2019 127 14 06 49 PMT counter failure | 1 | 1 | B12719.191 | CARE | #191 | 05:20:59 | | 1 | -1 |
Instrument constant changes for the week before 2019127 (1 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 32 rows)
HG burned out? (AP results when HG is ON) (0 sec, 77 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 191 | 2019127 | 2 | 4.4 |
Non-critical warnings from b-files' comments for days between 2019127 and 2019127 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 126 05 26 08 Reset Warm -> Tepid from 2019 126 05 26 03 until 2019 126 05 26 03 . Counts: 1,0 | 1 | 3 | B12719.191 | CARE | #191 | 05:20:08 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 4 | 5 | B12719.191 | CARE | #191 | 16:11:25 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | zs |
| 127 | CARE | #191 | ecnouv | 145 | 148 | 1 | 1 | 8 | 17 | 14 | 1 | 7 | 27 |
(0 sec)
New operators? (1 sec, 29 rows)
Relative and absolute humidity for day 2019127 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 11.7 | 1.9 |