Database last updated on Sun, 13 Sep 2026 13:40:55 +0000 by localhost
Critical warnings from b-files' comments for days between 2019065 and 2019065 (1 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 064 06 22 21 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.688570 Volts. | 12 | 3 | B06519.191 | CARE | #191 | 05:36:44 | | 1 | -1 |
| rl: | 2019 065 14 51 07 PMT counter failure | 1 | 1 | B06519.191 | CARE | #191 | 05:37:11 | | 1 | -1 |
Instrument constant changes for the week before 2019065 (2 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 23 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 191 | 2019065 | 2 | 8.8 |
Non-critical warnings from b-files' comments for days between 2019065 and 2019065 (1 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2019 064 05 43 49 Reset Warm -> Tepid from 2019 064 05 43 43 until 2019 064 05 43 44 . Counts: 1,0 | 1 | 3 | B06519.191 | CARE | #191 | 05:36:42 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | B06519.191 | CARE | #191 | 21:04:29 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | zs |
| 065 | CARE | #191 | ecnouv | 115 | 121 | 1 | 1 | 12 | 15 | 13 | 1 | 8 | 17 |
(1 sec)
Relative and absolute humidity for day 2019065 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 6.5 | 1.1 |