Database last updated on Tue, 15 Sep 2026 07:40:55 +0000 by localhost
Critical warnings from b-files' comments for days between 2018171 and 2018171 (1 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2018 170 05 27 28 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.640179 Volts. | 23 | 3 | B17118.191 | TORON | #191 | 05:30:56 | | 1 | -1 |
| rl: | 2018 171 16 17 35 PMT counter failure | 1 | 1 | B17118.191 | TORON | #191 | 05:31:28 | | 1 | -1 |
Instrument constant changes for the week before 2018171 (2 sec, 0 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 191 | 2018171 | 1 | 11.5 |
Non-critical warnings from b-files' comments for days between 2018171 and 2018171 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2018 170 05 26 26 Reset Warm -> Tepid from 2018 170 05 26 21 until 2018 170 05 26 21 . Counts: 1,0 | 1 | 3 | B17118.191 | TORON | #191 | 05:30:55 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 4 | 5 | B17118.191 | TORON | #191 | 16:40:40 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
| 171 | TORONTO | #191 | ecnouv | 137 | 141 | 1 | 1 | 15 | 4 | 26 | 23 | 1 | 5 | 42 |
(1 sec)
New operators? (0 sec, 59 rows)
Relative and absolute humidity for day 2018171 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 191 | 6.1 | 2.4 |