Database last updated on Mon, 02 Feb 2026 02:40:49 +0000 by localhost
Critical warnings from b-files' comments for days between 2023200 and 2023200 (0 sec, 7 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 200 14 42 58 Motor #3: appears stuck with a reference sensor activated. | 2 | 1 | B20023.190 | TORON | #190 | 05:29:30 | | 1 | -1 |
| rl: | 2023 200 15 31 40 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 7 | 3 | B20023.190 | TORON | #190 | 05:29:48 | | 1 | -1 |
| rl: | 2023 200 15 31 40 New value set for watermark LOOP.RUN.TIME: 699.000025 millisec. | 7 | 3 | B20023.190 | TORON | #190 | 05:29:48 | | 1 | -1 |
| pd: | Brewer failed to respond 5 times | 1 | 1 | B20023.190 | TORON | #190 | 14:31:18 | | 1 | -1 |
| hg: | Lamp not detected f1= 5 f4= 9 | 1 | 1 | B20023.190 | TORON | #190 | 15:13:27 | | 1 | -1 |
| hg: | Lamp not on... hg test terminated F1= 5 F4= 9 | 1 | 1 | B20023.190 | TORON | #190 | 15:13:27 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 8 | 5 | B20023.190 | TORON | #190 | 17:18:08 | | 1 | -1 |
Instrument constant changes for the week before 2023200 (7 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 16 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 190 | 2023200 | 1 | 22.2 |
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 190 | 2023200 | 2 | 79.9 |
Non-critical warnings from b-files' comments for days between 2023200 and 2023200 (0 sec, 6 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| rl: | 2023 200 14 42 58 Motor #3: appears stuck with a reference sensor activated. | 1 | 3 | B20023.190 | TORON | #190 | 05:29:30 | | 1 | -1 |
| rl: | 2023 200 14 42 58 Motor #11: appears stuck with a reference sensor activated. | 1 | 3 | B20023.190 | TORON | #190 | 05:29:32 | | 1 | -1 |
| rl: | 2023 200 15 08 07 Bad command string. Command string buffer cleared. | 1 | 3 | B20023.190 | TORON | #190 | 05:29:43 | | 1 | -1 |
| rl: | 2023 200 15 08 07 Bad command string. Command string buffer cleared. | 1 | 3 | B20023.190 | TORON | #190 | 05:29:43 | | 1 | -1 |
| rl: | 2023 200 15 08 07 < | 1 | 3 | B20023.190 | TORON | #190 | 05:29:45 | | 1 | -1 |
| si: | New SI, HC = | 1 | 99 | B20023.190 | TORON | #190 | 15:01:28 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | ed | hg | hp | sl | zs |
| 200 | TORONTO | #190 | 82 | 88 | 2 | 21 | 24 | 3 | 22 |
(1 sec)
Dead time (0 sec)
Nothing to report.
Standard lamp R6 (2 sec)
| File name | Count | Serial | Expected (+/-20) | Measured | 30-day history |
| B20023.190 | 3 | #190 | 408 | 386 |