Database last updated on Sun, 06 Sep 2026 15:11:08 +0000 by localhost

Critical warnings from b-files' comments for days between 2021104 and 2021104 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 103 13 40 57 New value set for watermark MOISTURE.HIGH: 12.400724 .93B10421.190TORON#19005:25:551-1
rl:2021 103 13 40 57 Motor #3: appears stuck with a reference sensor activated.21B10421.190TORON#19005:25:581-1
rl:2021 104 11 12 41 New value set for watermark LOOP.RUN.TIME: 684.000015 millisec.73B10421.190TORON#19005:26:061-1
ds:Suppressed oscillation of filter wheel.145B10421.190TORON#19013:22:251-1

Instrument constant changes for the week before 2021104 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
190202110425.8

Non-critical warnings from b-files' comments for days between 2021104 and 2021104 (0 sec, 7 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2021 103 13 40 56 Reset Tepid -> Tepid from 2021 103 13 34 07 until 2021 103 13 40 51 . Counts: 1,013B10421.190TORON#19005:25:521-1
rl:2021 103 13 40 57 New value set for watermark MOISTURE.HIGH: 12.400724 .23B10421.190TORON#19005:25:551-1
rl:2021 103 13 40 57 Motor #3: appears stuck with a reference sensor activated.13B10421.190TORON#19005:25:581-1
rl:2021 103 13 40 57 Motor #11: appears stuck with a reference sensor activated.13B10421.190TORON#19005:26:001-1
hg:HOME key pressed199B10421.190TORON#19012:48:201-1
hg:Going to menu199B10421.190TORON#19012:48:341-1
sc:SC intensity too low ( 271/cy), skipping.13B10421.190TORON#19021:57:221-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfzhghpsluvzs
104TORONTO#190menu,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 93.06301)646921312251522
(1 sec)

Relative and absolute humidity for day 2021104 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
19011.72.2