| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| fr: | 11000 steps - QUIT - Micrometer #1 reference not found | 1 | 1 | B07820.190 | TORON | #190 | 05:42:41 | 1 | -1 | |
| hp: | Spectrometer alignment trouble. | 8 | 1 | B07820.190 | TORON | #190 | 05:47:12 | 1 | -1 | |
| re: | Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu. | 1 | 1 | B07820.190 | TORON | #190 | 06:43:57 | 1 | -1 | |
| sl: | Lamp not on... sl test terminated F1= 2 F4= 1 | 1 | 1 | B07820.190 | TORON | #190 | 13:50:33 | 1 | -1 | |
| hp: | Brewer failed to respond 5 times | 1 | 1 | B07820.190 | TORON | #190 | 14:05:42 | 1 | -1 | |
| rl: | 2020 078 14 06 08 Motor #11: appears stuck with a reference sensor activated. | 2 | 1 | B07820.190 | TORON | #190 | 14:41:18 | 1 | -1 | |
| rl: | 2020 078 14 21 12 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.618672 Volts. | 6 | 3 | B07820.190 | TORON | #190 | 14:41:22 | 1 | -1 |
| Serial | YYYYJJJ | HG large | max correction |
|---|---|---|---|
| 190 | 2020078 | 7 | 20.8 |
| Serial | YYYYJJJ | HP large | max correction |
|---|---|---|---|
| 190 | 2020078 | 1 | 25.8 |
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| re: | Going to menu | 4 | 99 | B07820.190 | TORON | #190 | 06:43:57 | 1 | -1 | |
| fr: | HOME key pressed | 3 | 99 | B07820.190 | TORON | #190 | 12:48:57 | 1 | -1 | |
| re: | Azimuth discrepancy = 5039 | 1 | 99 | B07820.190 | TORON | #190 | 14:08:49 | 5039 | -30 | 30 |
| rl: | 2020 078 14 06 07 Reset No -> Tepid from 2020 078 14 02 30 until 1992 078 14 06 01 . Counts: 1,0 | 1 | 3 | B07820.190 | TORON | #190 | 14:41:16 | 1 | -1 | |
| rl: | 2020 078 14 06 08 Motor #11: appears stuck with a reference sensor activated. | 1 | 3 | B07820.190 | TORON | #190 | 14:41:18 | 1 | -1 | |
| rl: | 2020 078 14 07 00 Reset Warm -> Warm from 2020 078 14 06 57 until 2020 078 14 06 57 . Counts: 1,1 | 1 | 3 | B07820.190 | TORON | #190 | 14:41:19 | 1 | -1 | |
| rl: | 2020 078 14 07 01 Motor #10: appears stuck with a reference sensor activated. | 1 | 3 | B07820.190 | TORON | #190 | 14:41:21 | 1 | -1 | |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | B07820.190 | TORON | #190 | 15:15:49 | 1 | -1 | |
| ds: | DS intensity too high ( 382517/cy) quitting. | 1 | 99 | B07820.190 | TORON | #190 | 15:17:10 | 1 | -1 |
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | hg | hp | rs | sl | uv | zs |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 078 | TORONTO | #190 | ecsc,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 93.015),Starting schedule calo3 | 98 | 98 | 1 | 1 | 20 | 16 | 1 | 5 | 2 | 17 |
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
|---|---|---|---|---|
| 190 | 30.0 | 3.0 |