Database last updated on Wed, 09 Sep 2026 11:40:56 +0000 by localhost

Critical warnings from b-files' comments for days between 2020065 and 2020065 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 065 07 57 33 New value set for watermark SUPPLY.VOLTAGE.HIGH[-5VSEC]: -4.699324 Volts.53B06520.190TORON#19005:32:041-1
ds:Suppressed oscillation of filter wheel.105B06520.190TORON#19013:29:511-1

Instrument constant changes for the week before 2020065 (5 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 26 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
1902020065218.9

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1902020065122.9

Non-critical warnings from b-files' comments for days between 2020065 and 2020065 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
rl:2020 064 05 35 06 Reset Warm -> Tepid from 2020 064 05 35 00 until 2020 064 05 35 01 . Counts: 1,013B06520.190TORON#19005:32:021-1
ux:HOME key pressed199B06520.190TORON#19016:10:461-1
ux:Going to menu199B06520.190TORON#19016:11:351-1
ds:DS intensity too high ( 255133/cy) quitting.299B06520.190TORON#19019:21:561-1
ds:DS intensity too low ( 134/cy), skipping.199B06520.190TORON#19021:57:181-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
065TORONTO#190ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.453)5354213031412254217
(0 sec)

Relative and absolute humidity for day 2020065 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1901.80.2