Database last updated on Fri, 13 Mar 2026 14:40:43 +0000 by localhost

Critical warnings from b-files' comments for days between 2026032 and 2026032 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.41B03226.187TORON#18713:44:431-1
hg:Lamp not on... hg test terminated F1= 0 F4= 741B03226.187TORON#18714:59:511-1
hg:Lamp not detected f1= 0 f4= 731B03226.187TORON#18714:59:521-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B03226.187TORON#18720:04:441-1

Instrument constant changes for the week before 2026032 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 5 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202603223.6

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720260323170.0

Non-critical warnings from b-files' comments for days between 2026032 and 2026032 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.15B03226.187TORON#18714:54:251-1
re:Azimuth discrepancy = -41199B03226.187TORON#18720:03:51-41-3030
pd:Going to menu199B03226.187TORON#18720:04:451-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtfmfzhghprssluvzs
032TORONTO#187Attempting recovery FR,ec3uv38381245104614198
(0 sec)

Relative and absolute humidity for day 2026032 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18710.81.1