Database last updated on Fri, 13 Mar 2026 19:17:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2026025 and 2026025 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 931B02526.187TORON#18717:08:071-1
hg:Lamp not detected f1= 0 f4= 931B02526.187TORON#18717:08:081-1

Instrument constant changes for the week before 2026025 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202602544.5

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720260251030.0

Non-critical warnings from b-files' comments for days between 2026025 and 2026025 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Azimuth discrepancy = -74199B02526.187TORON#18705:43:40-74-3030

Measurements and tests summary.


_jday_loc_serial_skcdtedfmfzhghprssluvzs
025TORONTO#187ec3uv,Next schedule will be ec3uv (now ec3uv ZS= 93 ZS= 93.153)232218162627347
(1 sec)

Relative and absolute humidity for day 2026025 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18721.41.8