Database last updated on Tue, 27 Jan 2026 20:10:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2025055 and 2025055 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 621B05525.187TORON#18706:05:041-1
hg:Lamp not detected f1= 0 f4= 621B05525.187TORON#18706:05:051-1
hg:HG failed11B05525.187TORON#18706:12:35failed1-1
re:Previous RE didn11B05525.187TORON#18706:13:571-1

Instrument constant changes for the week before 2025055 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 5 rows)

HG burned out? (AP results when HG is ON) (0 sec, 1 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1872025055379.9

Non-critical warnings from b-files' comments for days between 2025055 and 2025055 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
pd:Going to menu199B05525.187TORON#18706:13:581-1
ds:DS intensity too high ( 263428/cy) quitting.299B05525.187TORON#18717:15:471-1
ds:Suppressed oscillation of filter wheel.45B05525.187TORON#18717:16:011-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfzhghpslzs
055TORONTO#187ecnouv,Next schedule will be ecnouv (now ecnouv ZS= 93 ZS= 93.038)8081271218640
(1 sec)

Relative and absolute humidity for day 2025055 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18714.42.0