Database last updated on Sun, 01 Feb 2026 10:40:49 +0000 by localhost

Critical warnings from b-files' comments for days between 2023363 and 2023363 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 021B36323.187TORON#18701:17:101-1
hg:No intensity detected. Check PMT.11B36323.187TORON#18703:12:421-1
hp:Spectrometer alignment trouble.301B36323.187TORON#18705:39:271-1
sl:Lamp not on... sl test terminated F1= 0 F4= 2331B36323.187TORON#18714:20:541-1

Instrument constant changes for the week before 2023363 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 15 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202336364.8

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720233632779.9

Measurements and tests summary.


_jday_loc_serial_skcdtedfmhghprssluv
363TORONTO#187ecsc,Must do HG now,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.351)123346751276
(0 sec)

Relative and absolute humidity for day 2023363 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1877.00.8