Database last updated on Wed, 02 Sep 2026 01:41:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2023345 and 2023345 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11#18701:58:401-1
sl:Lamp not on... sl test terminated F1= 0 F4= 6421#18713:14:381-1
hp:Spectrometer alignment trouble.181#18713:16:071-1
hg:Lamp not detected f1= 0 f4= 8151#18713:48:001-1

Instrument constant changes for the week before 2023345 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 13 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202334537.3

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720233459280.0

Non-critical warnings from b-files' comments for days between 2023345 and 2023345 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
pd:Going to menu199#18701:58:411-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdthghprssluv
345TORONTO#187calo3,Must do HG now,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 94.075),Waiting until ZA =223461413229
(1 sec)

Relative and absolute humidity for day 2023345 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1877.40.8