Database last updated on Sun, 01 Feb 2026 19:40:55 +0000 by localhost

Critical warnings from b-files' comments for days between 2023342 and 2023342 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 30441#18712:39:161-1
hg:Lamp not detected f1= 0 f4= 1741#18712:48:151-1
hp:Spectrometer alignment trouble.371#18713:09:051-1
hg:No intensity detected. Check PMT.21#18722:06:201-1

Instrument constant changes for the week before 2023342 (2 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 16 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
1872023342183.1

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720233423979.9

Measurements and tests summary.


_jday_loc_serial_skcdtedhghprssluv
342TORONTO#187calo3,Must do HG now,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 93.21601),Waiting until ZA =2283902142
(1 sec)

Relative and absolute humidity for day 2023342 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1877.81.2