Database last updated on Sun, 01 Feb 2026 22:40:50 +0000 by localhost

Critical warnings from b-files' comments for days between 2023182 and 2023182 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 1 f4= 011#18701:43:281-1
sl:Lamp not on... sl test terminated F1= 0 F4= 8231#18715:33:141-1
hp:Spectrometer alignment trouble.101#18716:43:391-1
fr:11000 steps51#18718:19:351-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.21#18718:34:131-1

Instrument constant changes for the week before 2023182 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

HG burned out? (AP results when HG is ON) (0 sec, 13 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202318223.3

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1872023182979.9

Non-critical warnings from b-files' comments for days between 2023182 and 2023182 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.45#18714:46:041-1
ds:DS intensity too high ( 384423/cy) quitting.299#18715:55:371-1
pd:Going to menu299#18718:34:141-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdthghprssluvzs
182TORONTO#187Attempting recovery FR,calo3,Going to menu,Must do HG now55572172323154
(0 sec)

Relative and absolute humidity for day 2023182 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18710.45.2