Database last updated on Tue, 03 Feb 2026 17:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2023176 and 2023176 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 0 F4= 21151#18718:12:241-1
hp:Spectrometer alignment trouble.151#18718:13:541-1

Instrument constant changes for the week before 2023176 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

HG burned out? (AP results when HG is ON) (0 sec, 4 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
187202317693.2

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
18720231762379.9

Non-critical warnings from b-files' comments for days between 2023176 and 2023176 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.45#18713:15:491-1
ds:DS intensity too high ( 225978/cy) quitting.199#18714:51:451-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
176TORONTO#187calo3,Must do HG now,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 93.032)858522126405624319
(0 sec)

Relative and absolute humidity for day 2023176 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18710.04.3