Database last updated on Mon, 14 Sep 2026 01:11:35 +0000 by localhost

Critical warnings from b-files' comments for days between 2019141 and 2019141 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 6 F4= 1111B14119.187CARE#18715:11:551-1
hg:Lamp not detected f1= 6 f4= 1111B14119.187CARE#18715:11:561-1
hg:HG cal too far off limits - resetting micrometer(s)11B14119.187CARE#18716:10:361-1
fr:11000 steps - QUIT - Micrometer #2 reference not found41B14119.187CARE#18717:39:091-1
pd:Brewer failed to respond 5 times21B14119.187CARE#18717:43:071-1

Instrument constant changes for the week before 2019141 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
1872019141401018.0

Non-critical warnings from b-files' comments for days between 2019141 and 2019141 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Zenith discrepancy = -1494199B14119.187CARE#18715:05:55-1494-1010
re:Azimuth discrepancy = -9938499B14119.187CARE#18715:07:10-9938-3030
hh:Bad or missing .cfg file creating new defaults.15B14119.187CARE#18717:01:171-1
hp:Going to menu199B14119.187CARE#18718:31:261-1
ds:Suppressed oscillation of filter wheel.35B14119.187CARE#18719:44:421-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedhghpslumuvzs
141CARE#187menu,S60MK32020123172122220
(0 sec)

Relative and absolute humidity for day 2019141 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18750.46.5