Database last updated on Sun, 13 Sep 2026 17:11:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2018247 and 2018247 (1 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hp:Spectrometer alignment trouble.41B24718.187TORON#18700:12:181-1
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B24718.187TORON#18700:33:341-1
hg:Lamp not on... hg test terminated F1= 0 F4= 12111B24718.187TORON#18712:46:341-1
hg:Lamp not detected f1= 0 f4= 12111B24718.187TORON#18712:46:351-1
ds:Suppressed oscillation of filter wheel.85B24718.187TORON#18715:28:291-1
hg:HG cal too far off limits - resetting micrometer(s)11B24718.187TORON#18722:48:261-1

Instrument constant changes for the week before 2018247 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
1872018247982.0

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1872018247779.9

Non-critical warnings from b-files' comments for days between 2018247 and 2018247 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-971 (+CAL)= 47 Now set to o3 offset = 7505199B24718.187TORON#18700:17:564710009000
fm:HOME key pressed299B24718.187TORON#18712:14:501-1
fm:Going to menu299B24718.187TORON#18712:14:531-1
ds:DS intensity too low ( 1/cy), skipping.199B24718.187TORON#18723:13:451-1

Measurements and tests summary.


_jday_loc_serial_skcdshghpslumuvzs
247TORONTO#187S60MK3412845563016
(1 sec)

New operators? (0 sec, 66 rows)

OperatorSerial
05187
19187

Relative and absolute humidity for day 2018247 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
18775.232.9