| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| hp: | Spectrometer alignment trouble. | 4 | 1 | B24718.187 | TORON | #187 | 00:12:18 | 1 | -1 | |
| re: | Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu. | 1 | 1 | B24718.187 | TORON | #187 | 00:33:34 | 1 | -1 | |
| hg: | Lamp not on... hg test terminated F1= 0 F4= 12 | 11 | 1 | B24718.187 | TORON | #187 | 12:46:34 | 1 | -1 | |
| hg: | Lamp not detected f1= 0 f4= 12 | 11 | 1 | B24718.187 | TORON | #187 | 12:46:35 | 1 | -1 | |
| ds: | Suppressed oscillation of filter wheel. | 8 | 5 | B24718.187 | TORON | #187 | 15:28:29 | 1 | -1 | |
| hg: | HG cal too far off limits - resetting micrometer(s) | 1 | 1 | B24718.187 | TORON | #187 | 22:48:26 | 1 | -1 |
| Serial | YYYYJJJ | HG large | max correction |
|---|---|---|---|
| 187 | 2018247 | 9 | 82.0 |
| Serial | YYYYJJJ | HP large | max correction |
|---|---|---|---|
| 187 | 2018247 | 7 | 79.9 |
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| fr: | Micrometer #1 reference found at step-971 (+CAL)= 47 Now set to o3 offset = 7505 | 1 | 99 | B24718.187 | TORON | #187 | 00:17:56 | 47 | 1000 | 9000 |
| fm: | HOME key pressed | 2 | 99 | B24718.187 | TORON | #187 | 12:14:50 | 1 | -1 | |
| fm: | Going to menu | 2 | 99 | B24718.187 | TORON | #187 | 12:14:53 | 1 | -1 | |
| ds: | DS intensity too low ( 1/cy), skipping. | 1 | 99 | B24718.187 | TORON | #187 | 23:13:45 | 1 | -1 |
| _jday | _loc | _serial | _skc | ds | hg | hp | sl | um | uv | zs |
|---|---|---|---|---|---|---|---|---|---|---|
| 247 | TORONTO | #187 | S60MK3 | 41 | 28 | 45 | 5 | 6 | 30 | 16 |
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
|---|---|---|---|---|
| 187 | 75.2 | 32.9 |