Database last updated on Sat, 24 Jan 2026 16:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2026018 and 2026018 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fz:Suppressed oscillation of filter wheel.95B01826.145TORON#14514:39:581-1
hg:Lamp not on... hg test terminated F1= 0 F4= 1311B01826.145TORON#14519:22:381-1
hg:Lamp not detected f1= 0 f4= 1311B01826.145TORON#14519:22:391-1

Instrument constant changes for the week before 2026018 (5 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
1452026018119.3

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1452026018670.0

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
018TORONTO#145ec3uv,Next schedule will be ec3uv (now ec3uv ZS= 93 ZS= 93.11)74742319152128338
(1 sec)

Relative and absolute humidity for day 2026018 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
1457.00.6