| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| hg: | Brewer failed to respond 5 times | 1 | 1 | B18525.145 | TORON | #145 | 11:00:52 | 1 | -1 | |
| hg: | Lamp not on... hg test terminated F1= 2 F4= 5 | 1 | 1 | B18525.145 | TORON | #145 | 11:57:41 | 1 | -1 | |
| hg: | Lamp not detected f1= 2 f4= 5 | 1 | 1 | B18525.145 | TORON | #145 | 11:57:42 | 1 | -1 | |
| ds: | Suppressed oscillation of filter wheel. | 15 | 5 | B18525.145 | TORON | #145 | 12:18:15 | 1 | -1 |
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| re: | Azimuth discrepancy = -3198 | 1 | 99 | B18525.145 | TORON | #145 | 11:51:23 | -3198 | -30 | 30 |
| fr: | Micrometer #1 reference found at step 8845 (+CAL)= 9811 Now set to o3 offset = 8965 | 1 | 99 | B18525.145 | TORON | #145 | 11:59:14 | 9811 | 1000 | 9000 |
| ds: | DS intensity too high ( 213030/cy) quitting. | 1 | 99 | B18525.145 | TORON | #145 | 17:04:25 | 1 | -1 |
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | fz | hg | hp | rs | sl | zs |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 185 | TORONTO | #145 | ecnouv,Next schedule will be ecnouv (now ecnouv ZS= 93 ZS= 93.013) | 160 | 161 | 2 | 2 | 11 | 5 | 19 | 24 | 2 | 6 | 32 |
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
|---|---|---|---|---|
| 145 | 7.4 | 3.4 |