Database last updated on Wed, 04 Feb 2026 21:10:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2022155 and 2022155 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 1 F4= 25371B15522.145TORON#14509:44:211-1
hg:Lamp not detected f1= 0 f4= 1341B15522.145TORON#14509:52:081-1
ds:Suppressed oscillation of filter wheel.445B15522.145TORON#14517:29:511-1

Instrument constant changes for the week before 2022155 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
145202215533.0

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
14520221551079.9

Non-critical warnings from b-files' comments for days between 2022155 and 2022155 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 380864/cy) quitting.199B15522.145TORON#14519:49:071-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhghprssluvzs
155TORONTO#145calo3,Must do HG now,Next schedule will be calo3 (now calo3 ZS= 93 ZS= 93.15),Waiting until ZA = -93 now -100.6,Waiting until ZA = -93 now -101.8,Waiting until ZA = -93 now -103,Waiting until ZA = -93 now -104.2,Waiting until ZA = -93 now -93.7,Waiting until ZA = -93 now -95.2,Waiting until ZA = -93 now -96.6,Waiting until ZA = -93 now -98,Waiting until ZA = -93 now -99.3818212125165152113
(1 sec)

Relative and absolute humidity for day 2022155 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
14516.05.1