Database last updated on Thu, 10 Sep 2026 03:41:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2020062 and 2020062 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sl:Lamp not on... sl test terminated F1= 1 F4= 12921B06220.145CARE#14514:21:581-1
hg:Lamp not detected f1= 2 f4= 1511B06220.145CARE#14514:29:211-1

Instrument constant changes for the week before 2020062 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 26 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
145202006233.0

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
1452020062280.0

Non-critical warnings from b-files' comments for days between 2020062 and 2020062 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
dp:Doing dispersion with internal HG.199B06220.145CARE#14506:39:491-1
fr:Micrometer #1 reference found at step 8899 (+CAL)= 9867 Now set to o3 offset = 8946199B06220.145CARE#14514:33:02986710009000
ux:HOME key pressed199B06220.145CARE#14514:56:231-1
ux:UV extended scan aborted13B06220.145CARE#14514:56:251-1
ux:Going to menu199B06220.145CARE#14514:56:371-1
ds:Suppressed oscillation of filter wheel.15B06220.145CARE#14519:34:021-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhghprssluvzs
062CARE#145ecsc3131213011416243821
(0 sec)

Relative and absolute humidity for day 2020062 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
14511.61.7