Database last updated on Thu, 05 Feb 2026 20:10:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2022098 and 2022098 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 3/cy), skipping.113B09822.085TORON#08512:12:091-1
hp:Spectrometer alignment trouble.31B09822.085TORON#08512:13:481-1
sl:Lamp not on... sl test terminated F1= 0 F4= 7541B09822.085TORON#08516:15:341-1

Instrument constant changes for the week before 2022098 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 1 rows)

HP with step corrections larger than 20 steps (0 sec, 1 rows)

SerialYYYYJJJHP largemax correction
08520220982679.9

Non-critical warnings from b-files' comments for days between 2022098 and 2022098 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ab_sk:HOME key pressed399B09822.085TORON#08510:26:211-1
pd:Going to menu299B09822.085TORON#08511:13:261-1
ds:DS intensity too high ( 353666/cy) quitting.399B09822.085TORON#08516:51:171-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedfmhghpsluvzs
098TORONTO#085calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.85501)5457224325651313
(1 sec)

Dead time (0 sec)

Nothing to report.

Standard lamp R6 (2 sec)

File nameCountSerialExpected (+/-20) Measured 30-day history
B09822.0855#085420387