Database last updated on Sun, 13 Sep 2026 12:41:32 +0000 by localhost
Critical warnings from b-files' comments for days between 2018318 and 2018318 (1 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| sl: | Lamp not on... sl test terminated F1= 0 F4= 5 | 3 | 1 | B31818.085 | TORON | #085 | 12:12:16 | | 1 | -1 |
| hp: | Spectrometer alignment trouble. | 2 | 1 | B31818.085 | TORON | #085 | 12:13:42 | | 1 | -1 |
Instrument constant changes for the week before 2018318 (1 sec, 0 rows)
HP with step corrections larger than 20 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HP large | max correction |
| 085 | 2018318 | 5 | 79.9 |
Non-critical warnings from b-files' comments for days between 2018318 and 2018318 (1 sec, 4 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| ds: | HOME key pressed | 1 | 99 | B31818.085 | TORON | #085 | 16:10:44 | | 1 | -1 |
| ds: | Going to menu | 1 | 99 | B31818.085 | TORON | #085 | 16:10:48 | | 1 | -1 |
| ds: | DS intensity too high ( 251372/cy) quitting. | 2 | 99 | B31818.085 | TORON | #085 | 17:38:49 | | 1 | -1 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | B31818.085 | TORON | #085 | 17:44:39 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | hp | rs | sl | uv | zs |
| 318 | TORONTO | #085 | ecsc | 40 | 40 | 2 | 1 | 1 | 18 | 24 | 2 | 5 | 43 | 23 |
(1 sec)
New operators? (0 sec, 192 rows)
Dead time (1 sec)
| File name | Count | Serial | Expected (+/-5) | Measured | 30-day history |
| B31818.085 | 2 | #085 | 28 | 32 |