Database last updated on Sun, 13 Sep 2026 17:11:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2018264 and 2018264 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 1 f4= 1411B26418.084TORON#08406:35:111-1
hg:Lamp not on... hg test terminated F1= 1 F4= 1411B26418.084TORON#08406:35:111-1

Instrument constant changes for the week before 2018264 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0842018264492.0

Non-critical warnings from b-files' comments for days between 2018264 and 2018264 (1 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.25B26418.084TORON#08416:51:381-1
sc:SC intensity too low ( 205/cy), skipping.13B26418.084TORON#08420:53:551-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
264TORONTO#084calsc1051062129833282610
(1 sec)

New operators? (0 sec, 30 rows)

Relative and absolute humidity for day 2018264 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
08410.84.6