Database last updated on Mon, 14 Sep 2026 00:11:06 +0000 by localhost

Critical warnings from b-files' comments for days between 2018260 and 2018260 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 1531B26018.084TORON#08406:14:461-1
hg:Lamp not detected f1= 0 f4= 1531B26018.084TORON#08406:14:471-1
hg:HG failed11B26018.084TORON#08406:48:42failed1-1

Instrument constant changes for the week before 2018260 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (1 sec, 3 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0842018260492.0

Non-critical warnings from b-files' comments for days between 2018260 and 2018260 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:Suppressed oscillation of filter wheel.25B26018.084TORON#08413:28:461-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
260TORONTO#084calsc110114211483227267
(0 sec)

New operators? (0 sec, 95 rows)

Relative and absolute humidity for day 2018260 (3 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0849.84.1