Database last updated on Sun, 01 Feb 2026 18:10:53 +0000 by localhost
Critical warnings from b-files' comments for days between 2024043 and 2024043 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hg: | Lamp not on... hg test terminated F1= 2 F4= 384 | 2 | 1 | | | #069 | 18:23:39 | | 1 | -1 |
| hg: | Lamp not detected f1= 2 f4= 384 | 2 | 1 | | | #069 | 18:23:40 | | 1 | -1 |
| hg: | HG failed | 1 | 1 | | | #069 | 18:27:38 | failed | 1 | -1 |
Instrument constant changes for the week before 2024043 (3 sec, 0 rows)
HG burned out? (AP results when HG is ON) (0 sec, 15 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 069 | 2024043 | 6 | 17.3 |
Non-critical warnings from b-files' comments for days between 2024043 and 2024043 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| uv: | HOME key pressed | 1 | 99 | | | #069 | 19:40:04 | | 1 | -1 |
| uv: | UV standard scan aborted | 1 | 3 | | | #069 | 19:40:06 | | 1 | -1 |
| pd: | Going to menu | 1 | 99 | | | #069 | 19:40:21 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | dt | ed | hg | rs | sl | uv |
| 043 | EUREKA | #069 | 2 | 2 | 24 | 2 | 7 | 19 |
(0 sec)
Relative and absolute humidity for day 2024043 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 069 | 3.4 | 0.4 |