Database last updated on Fri, 11 Sep 2026 20:11:01 +0000 by localhost

Critical warnings from b-files' comments for days between 2020012 and 2020012 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 2 F4= 33221B01220.069EUREK#06906:14:311-1
hg:Lamp not detected f1= 2 f4= 33221B01220.069EUREK#06906:14:321-1
hg:HG failed11B01220.069EUREK#06906:14:33failed1-1

Instrument constant changes for the week before 2020012 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 23 rows)

HG burned out? (AP results when HG is ON) (0 sec, 12 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
069202001223.6

Non-critical warnings from b-files' comments for days between 2020012 and 2020012 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
w1:HOME key pressed199B01220.069EUREK#06919:10:061-1
w1:Going to menu199B01220.069EUREK#06919:10:071-1
sim:New SIM, HC = 2, NC = 745 Old = 2 , 745199B01220.069EUREK#06919:17:101-1

Measurements and tests summary.


_jday_loc_serial_skcdtedfmhgrssl
012EUREKA#069ecsc21611025
(0 sec)

Relative and absolute humidity for day 2020012 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0695.00.5