Database last updated on Tue, 08 Sep 2026 06:40:57 +0000 by localhost

Critical warnings from b-files' comments for days between 2020260 and 2020260 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 251B26020.055STONY#05507:55:561-1
hg:Lamp not detected f1= 0 f4= 251B26020.055STONY#05507:55:571-1
hg:HG failed31B26020.055STONY#05507:55:58failed1-1

Instrument constant changes for the week before 2020260 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0552020260311.9

Non-critical warnings from b-files' comments for days between 2020260 and 2020260 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 101/cy), skipping.499B26020.055STONY#05513:16:391-1
ds:Suppressed oscillation of filter wheel.45B26020.055STONY#05516:04:271-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
260STONY PLAIN#055ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.18801)858521319275322
(0 sec)

Relative and absolute humidity for day 2020260 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
05510.42.0