Database last updated on Wed, 09 Sep 2026 10:41:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2020252 and 2020252 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 2111B25220.055STONY#05508:03:361-1
hg:Lamp not on... hg test terminated F1= 0 F4= 2111B25220.055STONY#05508:03:361-1
hg:HG failed41B25220.055STONY#05508:03:37failed1-1

Instrument constant changes for the week before 2020252 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0552020252716.1

Non-critical warnings from b-files' comments for days between 2020252 and 2020252 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 213433/cy) quitting.399B25220.055STONY#05515:32:001-1
ds:Suppressed oscillation of filter wheel.25B25220.055STONY#05523:50:241-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
252STONY PLAIN#055ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.461)83852122323276116
(1 sec)

Relative and absolute humidity for day 2020252 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
05510.82.1