Database last updated on Wed, 09 Sep 2026 08:11:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2020237 and 2020237 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 1151B23720.055STONY#05508:02:481-1
hg:Lamp not detected f1= 0 f4= 1151B23720.055STONY#05508:02:491-1
hg:HG failed31B23720.055STONY#05508:02:50failed1-1

Instrument constant changes for the week before 2020237 (7 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
05520202371213.8

Non-critical warnings from b-files' comments for days between 2020237 and 2020237 (1 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 224657/cy) quitting.199B23720.055STONY#05516:09:531-1
ds:Suppressed oscillation of filter wheel.25B23720.055STONY#05519:58:141-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
237STONY PLAIN#055ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.125)858521127266621
(1 sec)

Relative and absolute humidity for day 2020237 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
05513.22.0