Database last updated on Wed, 04 Feb 2026 11:10:52 +0000 by localhost

Critical warnings from b-files' comments for days between 2023172 and 2023172 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:HG failed91#04619:35:21failed1-1
hg:Lamp not on... hg test terminated F1= 0 F4= 5121#04620:44:411-1
hg:Lamp not detected f1= 0 f4= 5121#04620:44:421-1

Instrument constant changes for the week before 2023172 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 2 rows)

HG burned out? (AP results when HG is ON) (0 sec, 5 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
046202317247943.0

Non-critical warnings from b-files' comments for days between 2023172 and 2023172 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hf:Illegal command (hf) in schedule , ignoring199#04618:43:191-1
pd:Going to menu299#04620:23:231-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
172TORONTO#046ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.266)373722462244320
(0 sec)

Relative and absolute humidity for day 2023172 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0465.42.4