Database last updated on Fri, 11 Sep 2026 00:40:59 +0000 by localhost

Critical warnings from b-files' comments for days between 2019209 and 2019209 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 7 f4= 1141B20919.046TORON#04610:48:451-1
hg:Lamp not on... hg test terminated F1= 7 F4= 1141B20919.046TORON#04610:48:451-1
hg:HG failed21B20919.046TORON#04610:52:35failed1-1
hg:HG cal too far off limits - resetting micrometer(s)21B20919.046TORON#04615:58:531-1
hg:*** Micrometer reset attempted - HG cal still failed12B20919.046TORON#04619:12:041-1

Instrument constant changes for the week before 2019209 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
046201920979.1

Non-critical warnings from b-files' comments for days between 2019209 and 2019209 (1 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ed:Going to menu199B20919.046TORON#04605:24:281-1
ds:DS intensity too low ( 10/cy), skipping.399B20919.046TORON#04610:04:251-1
zs:ZS intensity too low ( 92/cy), skipping.499B20919.046TORON#04610:04:401-1
pdb1frhg:HOME key pressed199B20919.046TORON#04610:50:311-1
ds:DS intensity too high ( 335746/cy) quitting.199B20919.046TORON#04614:22:111-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtfzhgrssluvzs
209TORONTO#046ecsc,Waiting until ZA = 0 now -24.9,Waiting until ZA = 0 now -25.2,Waiting until ZA = 0 now -25.5,Waiting until ZA = 0 now -26.1,Waiting until ZA = 0 now -26.8,Waiting until ZA = 0 now -27.6,Waiting until ZA = 0 now -28.6,Waiting until ZA = 0 now -29.74545223424319
(1 sec)

Relative and absolute humidity for day 2019209 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
04649.812.8