Database last updated on Thu, 10 Sep 2026 18:11:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2019199 and 2019199 (1 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 38 F4= 4211B19919.046TORON#04619:50:431-1
hg:Lamp not detected f1= 38 f4= 4211B19919.046TORON#04619:50:441-1
hg:HG cal too far off limits - resetting micrometer(s)11B19919.046TORON#04620:01:281-1
hg:*** Micrometer reset attempted - HG cal still failed12B19919.046TORON#04620:14:121-1
hg:HG failed11B19919.046TORON#04620:14:12failed1-1

Instrument constant changes for the week before 2019199 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 32 rows)

HG burned out? (AP results when HG is ON) (0 sec, 10 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
046201919926.1

Non-critical warnings from b-files' comments for days between 2019199 and 2019199 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ed:Going to menu199B19919.046TORON#04605:24:181-1
zs:ZS intensity too low ( 217/cy), skipping.299B19919.046TORON#04610:04:411-1
ds:DS intensity too high ( 288668/cy) quitting.499B19919.046TORON#04612:57:161-1
ds:Suppressed oscillation of filter wheel.15B19919.046TORON#04616:05:541-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtfmhgrssluvzs
199TORONTO#046ecsc62652823234218
(1 sec)

Relative and absolute humidity for day 2019199 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
04610.44.7