Database last updated on Fri, 06 Feb 2026 17:10:54 +0000 by localhost

Critical warnings from b-files' comments for days between 2022036 and 2022036 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 30731B03622.039TORON#03913:41:381-1
hg:Lamp not detected f1= 0 f4= 30731B03622.039TORON#03913:41:391-1
hg:HG failed611B03622.039TORON#03916:51:43failed1-1
fr:11000 steps - QUIT - Micrometer #1 reference not found11B03622.039TORON#03923:12:081-1

Instrument constant changes for the week before 2022036 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
039202203626129.6

Non-critical warnings from b-files' comments for days between 2022036 and 2022036 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too low ( 1/cy), skipping.499B03622.039TORON#03912:27:151-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
036TORONTO#039ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 96.343)252522432026212
(0 sec)

Relative and absolute humidity for day 2022036 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0397.00.6