Database last updated on Sat, 07 Feb 2026 18:40:49 +0000 by localhost

Critical warnings from b-files' comments for days between 2021308 and 2021308 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 43281B30821.039TORON#03913:49:131-1
hg:Lamp not detected f1= 0 f4= 43281B30821.039TORON#03913:49:141-1
fr:11000 steps - QUIT - Micrometer #1 reference not found151B30821.039TORON#03913:51:221-1
hg:HG failed121B30821.039TORON#03913:53:38failed1-1

Instrument constant changes for the week before 2021308 (3 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2021308 and 2021308 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:HOME key pressed299B30821.039TORON#03915:09:121-1
hg:Going to menu299B30821.039TORON#03915:09:191-1
r:Illegal command ( r) in schedule , ignoring199B30821.039TORON#03917:39:141-1
hg:DEL key pressed199B30821.039TORON#03918:28:021-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
308TORONTO#039ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.837)1122418263222
(0 sec)

Relative and absolute humidity for day 2021308 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03932.04.3