Database last updated on Wed, 11 Feb 2026 06:40:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2021073 and 2021073 (1 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
td:Date changed from 13/03/21 to 14/03/2111B07321.039TORON#03912:25:481-1
hg:Lamp not on... hg test terminated F1= 0 F4= 36811B07321.039TORON#03914:40:121-1
hg:Lamp not detected f1= 0 f4= 36811B07321.039TORON#03914:40:131-1
hg:HG cal too far off limits - resetting micrometer(s)11B07321.039TORON#03915:52:321-1
hg:*** Micrometer reset attempted - HG cal still failed12B07321.039TORON#03917:36:251-1
hg:HG failed11B07321.039TORON#03917:36:26failed1-1

Instrument constant changes for the week before 2021073 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 4 rows)

SerialYYYYJJJSL CurrentSL Voltage
03920210670.0712.37
03920210680.0712.37
03920210721.6310.91
03920210731.6310.72

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
039202107320923.0

Non-critical warnings from b-files' comments for days between 2021073 and 2021073 (3 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ed:Going to menu199B07321.039TORON#03905:26:551-1
ds:DS intensity too low ( 110/cy), skipping.499B07321.039TORON#03913:04:341-1

Measurements and tests summary.


_jday_loc_serial_skcaodedshgsluvzs
073TORONTO#039ecsc,menu,Must do HG now,Waiting until ZA = 0 now -46.1,Waiting until ZA = 0 now -46.3,Waiting until ZA = 0 now -46.6,Waiting until ZA = 0 now -47,Waiting until ZA = 0 now -47.5,Waiting until ZA = 0 now -48.1,Waiting until ZA = 0 now -48.8,Waiting until ZA = 0 now -49.5,Waiting until ZA = 0 now -50.42020131136
(4 sec)

Relative and absolute humidity for day 2021073 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0397.40.8