Database last updated on Wed, 09 Sep 2026 00:11:10 +0000 by localhost

Critical warnings from b-files' comments for days between 2020235 and 2020235 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:HG cal too far off limits - resetting micrometer(s)11B23520.039TORON#03916:59:551-1
hg:Lamp not detected f1= 3 f4= 422151B23520.039TORON#03917:22:491-1
hg:Lamp not on... hg test terminated F1= 3 F4= 422151B23520.039TORON#03917:22:491-1
hg:HG failed1071B23520.039TORON#03918:52:16failed1-1

Instrument constant changes for the week before 2020235 (1 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2020235 and 2020235 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Zenith discrepancy = 52399B23520.039TORON#03905:32:4452-1010
uv:UV standard scan finished, incident DUV = 966.9082 mW/m^2199B23520.039TORON#03916:16:38966.9082-1450

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhgrssluvzs
235TORONTO#039ecsc,Waiting until ZA = 0 now -32.4,Waiting until ZA = 0 now -32.631312118233222
(1 sec)

Relative and absolute humidity for day 2020235 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03910.06.4