Database last updated on Sun, 13 Sep 2026 03:11:03 +0000 by localhost

Critical warnings from b-files' comments for days between 2019057 and 2019057 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 14211B05719.039TORON#03902:54:271-1
hg:Lamp not detected f1= 0 f4= 14211B05719.039TORON#03902:54:281-1

Instrument constant changes for the week before 2019057 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0392019057312.7

Non-critical warnings from b-files' comments for days between 2019057 and 2019057 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 251392/cy) quitting.499B05719.039TORON#03914:47:171-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
057TORONTO#039ecsc55582114215275624
(0 sec)

Relative and absolute humidity for day 2019057 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03912.41.2