Database last updated on Sun, 13 Sep 2026 13:40:55 +0000 by localhost

Critical warnings from b-files' comments for days between 2019016 and 2019016 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 36421B01619.039TORON#03920:54:091-1
hg:Lamp not on... hg test terminated F1= 0 F4= 36421B01619.039TORON#03920:54:091-1

Instrument constant changes for the week before 2019016 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 8 rows)

HG burned out? (AP results when HG is ON) (0 sec, 10 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
03920190161092.0

Non-critical warnings from b-files' comments for days between 2019016 and 2019016 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.15B01619.039TORON#03919:07:451-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
016TORONTO#039ecsc34392135621264430
(0 sec)

New operators? (0 sec, 71 rows)

Relative and absolute humidity for day 2019016 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03910.41.0