Database last updated on Sat, 31 Jan 2026 07:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2024190 and 2024190 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 15711#03108:00:551-1
hg:Lamp not detected f1= 0 f4= 15771#03108:00:561-1
hg:HG failed2821#03108:00:57failed1-1

Instrument constant changes for the week before 2024190 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
03120241904213.7

Non-critical warnings from b-files' comments for days between 2024190 and 2024190 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fz:Suppressed oscillation of filter wheel.15#03105:39:501-1
ds:DS intensity too low ( 80/cy), skipping.499#03108:30:011-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
190RESOLUTE,RESOLUTE#031ecsc,Must do HG now50502264223149
(0 sec)

Relative and absolute humidity for day 2024190 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03120.23.5