Database last updated on Mon, 02 Feb 2026 17:10:51 +0000 by localhost

Critical warnings from b-files' comments for days between 2023082 and 2023082 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Zenith ZEROING FAILURE12#03119:51:471-1
hg:Lamp not on... hg test terminated F1= 0 F4= 02621#03120:43:321-1
hg:No intensity detected. Check PMT.1251#03120:43:331-1
hg:Lamp not detected f1= 0 f4= 02621#03120:43:341-1
hg:HG failed1311#03121:07:17failed1-1

Instrument constant changes for the week before 2023082 (5 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 3 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
03120230824922.0

Non-critical warnings from b-files' comments for days between 2023082 and 2023082 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:HOME key pressed399#03120:06:091-1
pd:Going to menu399#03120:06:231-1

Measurements and tests summary.


_jday_loc_serial_skcdthgrssl
082RESOLUTE#031Attempting recovery FR,menu,Must do HG now1412
(0 sec)

Relative and absolute humidity for day 2023082 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03146.21.2