Database last updated on Mon, 02 Feb 2026 18:40:50 +0000 by localhost

Critical warnings from b-files' comments for days between 2023042 and 2023042 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 2811071#03107:00:361-1
hg:Lamp not detected f1= 0 f4= 2811071#03107:00:371-1
hg:HG failed531#03107:00:38failed1-1

Instrument constant changes for the week before 2023042 (5 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
031202304223.1

Non-critical warnings from b-files' comments for days between 2023042 and 2023042 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ab_sk:HOME key pressed299#03115:08:231-1
pd:Going to menu199#03115:22:171-1

Measurements and tests summary.


_jday_loc_serial_skcdtedhgrssluv
042RESOLUTE#031Attempting recovery RESET,ecsc,Next schedule will be ecsc (now ecsc ZS= 93 ZS= 93.042)173171244
(1 sec)

Relative and absolute humidity for day 2023042 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03116.02.1