Database last updated on Mon, 07 Sep 2026 09:11:03 +0000 by localhost

Critical warnings from b-files' comments for days between 2020357 and 2020357 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu.11B35720.031RESOL#03103:35:441-1
hg:*** Micrometer reset attempted - HG cal still failed12B35720.031RESOL#03108:13:081-1
hg:HG failed31B35720.031RESOL#03108:13:09failed1-1
hg:Lamp not detected f1= 0 f4= 3931B35720.031RESOL#03123:53:241-1
hg:Lamp not on... hg test terminated F1= 0 F4= 3931B35720.031RESOL#03123:53:241-1

Instrument constant changes for the week before 2020357 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0312020357511.5

Non-critical warnings from b-files' comments for days between 2020357 and 2020357 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Going to menu299B35720.031RESOL#03108:13:111-1

Measurements and tests summary.


_jday_loc_serial_skcfmhgsl
357RESOLUTE#031ecsc,Starting schedule ecsc751
(0 sec)

Relative and absolute humidity for day 2020357 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
03112.82.3