Database last updated on Sat, 12 Sep 2026 22:11:02 +0000 by localhost

Critical warnings from b-files' comments for days between 2019107 and 2019107 (0 sec, 6 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 0 F4= 0471B10719.026TORON#02605:35:451-1
hg:Lamp not detected f1= 0 f4= 0471B10719.026TORON#02605:35:461-1
hg:No intensity detected. Check PMT.71B10719.026TORON#02605:35:461-1
hg:HG failed211B10719.026TORON#02605:35:46failed1-1
hg:HG cal too far off limits - resetting micrometer(s)21B10719.026TORON#02610:25:081-1
fr:11000 steps - QUIT - Micrometer #1 reference not found61B10719.026TORON#02610:31:311-1

Instrument constant changes for the week before 2019107 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
02620191071829.0

Non-critical warnings from b-files' comments for days between 2019107 and 2019107 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-1162 (+CAL)=-243 Now set to o3 offset = 2936199B10719.026TORON#02606:05:39(+CAL)=-24310009000

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmhgrssluvzs
107TORONTO#026ecsc7777213244255522
(1 sec)

Relative and absolute humidity for day 2019107 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02613.63.4