Database last updated on Sat, 12 Sep 2026 18:41:08 +0000 by localhost
Critical warnings from b-files' comments for days between 2019105 and 2019105 (0 sec, 5 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| fr: | 11000 steps - QUIT - Micrometer #1 reference not found | 3 | 1 | B10519.026 | TORON | #026 | 05:37:34 | | 1 | -1 |
| hg: | Lamp not detected f1= 0 f4= 0 | 36 | 1 | B10519.026 | TORON | #026 | 05:39:39 | | 1 | -1 |
| hg: | No intensity detected. Check PMT. | 4 | 1 | B10519.026 | TORON | #026 | 05:39:39 | | 1 | -1 |
| hg: | Lamp not on... hg test terminated F1= 0 F4= 0 | 36 | 1 | B10519.026 | TORON | #026 | 05:39:39 | | 1 | -1 |
| hg: | HG failed | 13 | 1 | B10519.026 | TORON | #026 | 05:39:40 | failed | 1 | -1 |
Instrument constant changes for the week before 2019105 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 24 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 026 | 2019105 | 13 | 17.0 |
Non-critical warnings from b-files' comments for days between 2019105 and 2019105 (0 sec, 1 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| uv: | UV standard scan finished, incident DUV = 479.4648 mW/m^2 | 1 | 99 | B10519.026 | TORON | #026 | 18:15:48 | 479.4648 | -1 | 450 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fm | hg | rs | sl | uv | zs |
| 105 | TORONTO | #026 | ecsc | 79 | 79 | 2 | 1 | 24 | 33 | 2 | 6 | 57 | 30 |
(0 sec)
New operators? (0 sec, 62 rows)
| Operator | Serial |
| (CUT) | 026 |
| (HH:MM) | 026 |
Relative and absolute humidity for day 2019105 (2 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 026 | 14.8 | 2.3 |